Avante Infantia NB1 Infant Incubator
Avante Infantia NB1 Infant Incubator features servo-control, as well as quiet windows that minimize disturbance to the infant during nursing.
Avante Infantia NB1 Infant Incubator Features:
- Servo-controlled by micro-computer
- Quiet windows reduce disturbance to the baby during nursing
- Adjustable inclination of the bassinet
- Touch switch panel with LED digital display
- Set temperature, real temperature and auxiliary temperature are displayed separately
- Multiple self-checking alarms with audible and visible
- Heating power rate indicates 0 – 100%
- RS 232 connector
- Data output
- Standard Configuration:
- Hood
- Infant bed
- Temperature controller
- Skin temperature sensor
- Aluminum water tank
- Air filter
- Guardrail
- I.V. pole and cabinet
Avante Infantia NB1 Infant Incubator Specifications:
- Power Supply: AC110 – 240V/50-60HZ
- Power Input: 600VA
- Air Temperature Control Range: 25° C — 37° C
- Skin Temperature Control Range: 34° C — 37° C
- Sensor Precision: ≤0.3° C
- Changing of Temperature: ≤0.5° C
- Uniformity of Temperature: ≤0.8° C
- Temperature Rising Time: ≤30min
- Noise Level: ≤55dB (A)
- Inclination of Bassinet: 0° – 5°
- Alarming
- Audible and Visible Alarm: Indicates the real temperature and cuts off the power supply of heater (except for the low deviation alarm), meanwhile, the skin temperature
- indication window can indicate the alarm code. For Alarm 1 – 6, the relevant light is on. For Alarm 7, all alarming lights are on. All alarm sounds can be canceled except for
- the power failure alarm and the system alarm.
- Over-Temp Alarm: Sounds when the air temperature control is ≤38° C, and the skin temperature control is ≤40° C
- Upper Deviation Alarm: Sounds when the air temperature control is +3.0° C, and the skin temperature control is +1.0° C
- Lower Deviation Alarm: Sounds when the air temperature control is -3.0° C, and the skin temperature control is -1.0° C
- Sensor Alarm: Sounds when the sensor is disconnected, open circuit, short circuit or put on the wrong place
- Fan Alarm: Sounds when the fan is blocked, speed ≤1000rpm or stop working
- Power Failure Alarm: Sounds when there is no power or the power cord is disconnected
- System Alarm: Sounds when there is an E2ROM malfunction inside of temperature controller
- Gross Weight: Main body – 66 kg; Cabinet – 68 kg
- Measurement: Main body – 1,085 × 640 × 900 mm; Cabinet – 1,125 × 665 × 715 mm