Description
CareFusion Refurbished Alaris Pump & Syringe Module
CareFusion Refurbished Alaris Pump & Syringe Module increased use of advanced information technologies in the hospital environment promises improved access to and awareness of patient history and treatment plans at all levels of the patient care process. The skillful use of this data can provide an opportunity to improve an enterprise’s clinical, operational and financial performance.
Carefusion Alaris BD 8110 Features
The CareFusion Alaris 8110 Pump and Syringe module is intended for fluid delivery to a patient. The 8110 pump was designed to work with the Alaris PC unit and the Guardrails drug library system. The Alaris 8110 is a fluid delivery system, providing fluids, medications, blood, and blood products internally to a patient. The Syringe system uses a pressure sensing dispenser to reduce the infusion time and allow for precise infusions. Using the drug library system software accurate delivery of dosages can be delivered to the patient. The Alaris 8110 can be used for all patients including adults, pediatrics, and neonates.
- Intended for the infusion delivery of fluids, medications, blood and blood products
- Intravenous, intr-arterail and enteral used of delivering of fluids
- uses a pressure disc in the syringe system to to accurately monitor the line pressure and delivery
- Part of the Alaris Medley Point of Care System
- Guardrails drug library compatible
- lightweight – only 4.5 pounds
- Common User interface- same front panel keys as other modules
- Front loading of the Syringe
- Durable Design and can handle up to four additional modules attached to it
- Plunger driveled protected – extended handle to protect plunger
Carefusion Alaris BD 8110 Specifications
Bolus Volume, Maximum after Occlusion
Without Pressure Sensing Disc
- Pressure Setting Low: 0.512 Bolus Volume (mL)
- Pressure Setting Medium: 0.776 Bolus Volume (mL)
- Pressure Setting High: 1.103 Bolus Volume (mL)
With Pressure Sensing Disc
- Pressure Setting 300 mmHg: Back Off Disabled: 0.462; Back Off Enabled: 0.126
- Pressure Setting 500 mmHg: Back Off Disabled: 0.575; Back Off Enabled: 0.331
- Pressure Setting 1000 mmHg: Back Off Disabled: 0.839; Back Off Enabled: 0.323
Critical Volume
- Maximum over-infusion which can occur in the event of a single-fault condition will not exceed 2% of nominal syringe fill volume during loading and 1% of maximum syringe travel after syringe loading.
Flow Rate Range
- 0.01-9.99 ml/hr in 0.01 ml increments
- 10-99.9 ml/hr in 0.1 ml increments
- 100-999 ml/hr in 1 ml increments
Dimensions
- Width: 4.5″
- Height: 15″
- Depth: 7.5″
- Weight: 4.5 Lbs.
For more CareFusion Product click here.